Model : RN PET H5
Display screen : 5.5inch
Resolution : 1280*720 pixels
Type : Rechargeable lithium battery 4000 mAh 3.7V
Maximum charging time is less than 6 hours
Working time : 6 hours
Voltage : DC 5V 2A
Product Description Support OEM, reliable performance, cost-effective multi-parameter monitor solution
Features
- 5.5inch capacitive touch screen with friend Ul design
- Suitable for small and large animals
- Anti-defibrillator, anti-ESU , with Titration function
- Advanced blood oxygen technology, anti-motion interference
- Smart system with built-in manual
- Multiple configurations to choose:
WSpet H5:SPO2+NIBP+TEMP
wSpet H5S:SPO2
WSpet H5ST:SPO2+TEMP
WSpet H5N:NIBP+TEMP
WSpet H5P:SPO2+NIBP+ECG+TEMP - Optional: Wifi , wireless printer, Etco2 , carry bag
- Remote data transfer available, supporting to lOS/Android system (App)
- Rechargeable lion-battery backup time more than 6 hours continuously working
- Ultra-thin design, lightweight,compact easy to carry

Technology specifications
Product name | RN PET H5 |
Temperature | Technique: Thermistor probe (2.25K/10k); |
Measuring and Alarm Range: 0.0℃~50℃; | |
Unit: Celsius (℃); | |
Resolution: 0.1℃ or 1F; | |
Accuracy: ±0.1℃ (exclusive probe); | |
Pulse Rate | Measuring and Alarm Range: 25~250bpm; |
Accuracy: ±3bpm; | |
Resolution: 1bpm; | |
ECG | Lead mode: 3 Leads I,II,III; |
Gain: 1/4 file (X0.25) 2.5mm/mV; 1/2 file (X0.50) 5.0mm/mV; 1 file (X1) 10mm/mV; 2 files (X2) 20mm/mV; | |
Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s; | |
Baseline Recovery: < 3s after defibrillation (Monitor mode), < 1s after defibrillation (Surgery mode); | |
Bandwidth: Diagnostic 0.05~100, Monitor 0.5~40Hz, Surgery 1~20Hz; | |
NIBP | Method: Pulse wave oscillation method; |
Measure mode: Manual, Auto, Continue; | |
Measure interval in auto mode: 1 file (X1) 10mm/mV; 2 files (X2) 20mm/mV; 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480min; | |
STAT mode cycle time: Keep 5 minutes; | |
Measure and Alarm Range: Large Animal: SYS 30~280mmHg DIA 10~220mmHg MEAN 20~240mmHg, Small Animal: SYS 30~220mmHg DIA 10~160mmHg MEAN 20~170mmHg, Static pressure range: 0~300mmHg, Static pressure accuracy: 3mmHg; |
|
Maximum Mean error: 5mmHg; | |
Maximum Standard deviation: 8mmHg; | |
Static pressure accuracy: ±3mmHg; | |
Resolution: 1mmHg; | |
Accuracy: Maximum Mean error ± 5mmHg, Maximum Standard deviation<8mmHg; | |
Over pressure Protection: Large animal~300mmHg, Small animal~240mmHg; | |
Heart Rate | Measure range: Large animal 15~300 bpm, Small animal 15~350 bpm; |
Accuracy: 1% or 1bpm, whichever is greater; | |
Resolution: 1 bpm; | |
SpO2 | Measurement Range: 0~100%; |
Resolution: 1%; | |
Response Modes: Low, Medium, High; | |
Accuracy: 70%~100%: 2% <69%: unspecified; | |
EtCO2 | Measure method: Non-dispersive infrared (NDIR); |
Measure Range: 0~19.7% (0~150mmHg) 0~20kPa; | |
Resolution: 0.1 mmHg; | |
CO2 Accuracy: 0~40 mmHg, ±2 mmHg, 41~70 mmHg, ±5% of reading, 71~100 mmHg, ±8% of reading, 101~150 mmHg, ±10% of reading, at 760mmHg, ambient temperature of 25℃); |
|
Respiratory Rate: Range:3-150BrPM, Accuracy:±1bpm; | |
Transportation and Storage | Temperature: -20~50℃; |
Humidity: 10%~90%(non-condensing); | |
Atmosphere pressure: 86KPa~106KPa; | |
Respiration | Method:lmpedance between RA-LL; |
Gain: x0.25, x0.50, x1, x2, x4; | |
Respiration Rate:0~120/0-150BrPM; | |
Sweep speed: 6.25 mm/s,12.5 mm/s,25mm/s; | |
Resolution:1 BrPM; | |
Apnea Alarm:10~40s; | |
Measure range: Large animal: 0-120 bpm; | |
Small animal:0-150 bpm; | |
Trend & Reviewing | NlBP measurement reviewing: 1500groups; |
Waveform review: 3 hours 600groups; | |
Alarm event: 1000 groups of parameter alarms events and associated parameter; |